VA Series
Vapor Actuated Dial Thermometer
INSTRUCTION
SHEET
M1982/0902
General Description
The OMEGA
®
VA Series Vapor Actuated Thermometer is ideal for remote
readings, such as panel installations, with capillary lengths from 5 to 100
feet. This thermometer gives excellent readings above and below ambient
temperature and, being vapor actuated, is not subject to reading errors due
to either RFI/EMI interference or ambient temperature variations along the
capillary tube length.
Available Models
Front Flange U-Clamp (Panel Mount) Rear Flange
(Panel Mount) (Surface Mounted)
Figure 1. Case Style
Installing the Vapor Actuated Thermometer
Unpacking Instructions
Remove the Packing List and verify that you have received all equipment,
including the following (quantities in parentheses):
Vapor Actuated Dial Thermometer (1)
Operator’s Manual (1)
If you have any questions about the shipment, please call the OMEGA
Customer Service Department.
When you receive the shipment, inspect the container and equipment for
signs of damage. Note any evidence of rough handling in transit.
Immediately report any damage to the shipping agent.
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - VA Series

VA SeriesVapor Actuated Dial ThermometerINSTRUCTIONSHEETM1982/0902General DescriptionThe OMEGA®VA Series Vapor Actuated Thermometer is ideal for remot

Page 2

VA Series Vapor Actuated Dial ThermometerThe carrier will not honor damage claims unless all shippingmaterial is saved for inspection. After examining

Page 3

VA Series Vapor Actuated Dial ThermometerFlush Mount (Front Flange and U-Clamp)1. Make a circular cutout in the panel slightly larger than the case.2

Page 4

VA Series Vapor Actuated Dial ThermometerStandard Bulb Dimensions (inches)Immerse the thermometer stem at least A inches (or entire stem,whichever is

Page 5 - Operation

VA Series Vapor Actuated Dial ThermometerUse a thermowell in all pressurized applications. Be surecase temperature does not exceed 200°F (93°C).Thermo

Page 6 - Recalibration

VA Series Vapor Actuated Dial ThermometerVapor actuated thermometers give excellent readings when the bulbtemperature is either above or below the tem

Page 7 - Specifications

VA Series Vapor Actuated Dial ThermometerSpecificationsDial Sixes: 2", 2 1⁄2", 3 1⁄2", 4 1⁄2"Dial: Aluminum with white finish and

Page 8 - WARRANTY/DISCLAIMER

M1982/0901It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification